CLINICAL GUIDANCE | SECTION ONE
13
Automatic Gain Control (AGC) – front-end amplifier that reduces the
gain of high amplitude inputs to avoid distortion/peak clipping. The AGC
threshold depends upon the microphone sensitivity setting.
Autosensitivity – this algorithm automatically adjusts the microphone
sensitivity to lessen the effects of background noise.
1.3.1 Input Processing/SmartSound™ Options
ADRO
®
– Adaptive Dynamic Range Optimization - a digital pre-processing
signal algorithm designed to improve audibility of low-level sounds and
reduce the gain on higher level sounds to keep the signal level at the
desired point below C-level. This is available for the SPrint
, Freedom
, and
CP810 Sound Processors.
Whisper
a fast-acting compression circuit that makes soft sounds easier
to hear. This is available for the ESPrit™ 3G, Freedom, and CP810 Sound
Processors.
Beam
a beamforming algorithm that uses spatial input processing and
intelligent noise cancelling to automatically adjust microphone directionality
depending on the presence and type of noise sources.
It improves the signal-to-noise ratio (SNR) by reducing the loudness of
sound that originates from behind and to the sides of the listener and makes
it easier to hear sounds originating from the front. It is adaptive, so if the
noise source is moving dynamically, so too does the direction of maximum
attenuation. This is available for the Freedom and CP810 Sound Processors.
Zoomuses an algorithm that has a fixed directional pattern, where the
pattern does not change as a function of changes in spatial separation
between speech and noise, keeping the soundscape relatively constant.
It gives optimal blocking of diffuse noise, or when noise comes from
many directions simultaneously and the signal of interest is in front of the
listener. This is available for the CP810 Sound Processor.
Autosensitivity
– automatically adjusts microphone sensitivity based on
the noise floor of the surrounding environment.
The Custom Sound clinical programming software provides combinations
of these input processing options.