Cochlear SPrint Sound Processor Información técnica Pagina 21

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 78
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 20
CLINICAL GUIDANCE | SECTION ONE
19
1.5.3 Variable and Pseudomonopolar (PSMP) Stimulation
Variable and pseudomonopolar stimulation modes combine different BP
modes in the same program or MAP. For example, in variable mode some
channels may be programmed in BP1 and others in BP2, BP3, or BPx. Variable
mode can be used to increase the number of active channels in situations
where there are a reduced number of electrodes available due to electrode
anomalies or non-auditory sensations. Variable mode can also be used to
create wider stimulation modes for channels that have too high C-levels in the
nominal mode.
Pseudomonopolar is a specific application of variable mode. Typically it is
used in cases of partial insertion of the electrode array where the monopolar
reference electrodes are not available, e.g. in the case of the Nucleus 22
System. In these cases, pseudomonopolar is used to maximize the number
of active channels. In pseudomonopolar, when each of the intracochlear
electrodes is designated active, the indifferent electrode is an electrode that
is just outside the cochlea. In pseudomonopolar the current flow is reversed
because the indifferent electrode is basal to the active electrode. For example,
if electrodes 10 – 22 are inside the cochlea, each of these electrodes could be
referenced to electrode 9 that would be at the cochleostomy. Then channel
22 would have the widest BP mode (BP-12) and channel 10 would have the
narrowest mode (BP).
1.5.4 Common Ground (CG) Stimulation
In CG stimulation, current flows between the active electrode and all the other
electrodes on the array, which are connected together electronically to form
a single indifferent or reference (see Figure 4). The extracochlear electrodes
(MP1 and MP2) are not used. CG is an important mode because it can detect
electrode anomalies. As each electrode on the array is activated, the remaining
electrodes form the reference. Impedance telemetry uses CG to detect
potential electrode problems on the intracochlear array.
Just like MP modes, CG allows all 22 electrodes to be used as active compared
to BP where the maximum number of channels is 21. Due to the wider spread
of current in CG, this mode has been considered somewhat analogous to MP or
a wide BP mode. However, T- and C-levels in CG are typically higher than in MP.
Although CG is a programming option with the SPEAK strategy, MP modes are
still the recommendedrst choice. CG should not be used as a programming
mode with implant recipients who have partial insertions of the electrode array.
Because current is flowing to all electrodes on the array, it willow to any
electrodes outside the cochlea and this may produce non-auditory sensations.
Vista de pagina 20
1 2 ... 16 17 18 19 20 21 22 23 24 25 26 ... 77 78

Comentarios a estos manuales

Sin comentarios